| Pressure Specifications |
| Psi Range, Resolution |
-12.000 to 300.000 psi |
| Bar Range, Resolution |
-0.8273 to 20.6843 bar |
| kPa Range, Resolution |
-82.73 to 2068.43 kPa |
| One-Year Specification |
0.02% of full scale |
| Control Specification |
0.005% full scale minimum |
| Ranges |
| mA Measure, Source and Simulate |
0 to 24 mA |
| Volts DC Measure |
0 to 30 V DC |
| Resolution |
| mA DC Source, Simulate and Measure |
1 μA |
| Voltage DC Measurement |
1 mV |
| Accuracy |
0.01% ± 2 LSD all ranges (at 23°C ± 5°C) |
| Stability |
20 ppm of full scale /°C from -10 to +18°C and 28 to +50°C |
| Loop Compliance Voltage |
24 V DC at 20 mA |
| mA Simulate External Voltage Requirement |
12 to 30 V DC |
| Temperature |
0.01°F (0.01°C) |
| Drive Capability |
1200 Ω without HART resistor, 950 Ω with internal HART resistor |
| General Specifications |
| Temperature |
Operating: 14 to 122°F (-10 to 50°C) for measurement, 32 to 122°F (0 to 50°C) for pressure control
Storage: -4 to 140° (-20 to 60 °C) |
| Altitude |
Operating: 9842.52′ (3000 m)
Storage: 4265.092′ (1300 m) |